On-demand wireless infusion rate control in an implantable micropump for patient-tailored treatment of chronic conditions.
On-demand wireless infusion rate control in an implantable micropump for patient-tailored treatment of chronic conditions.
复制标题
植入式微型泵中的按需无线输注速率控制,用于针对患者定制的慢性病治疗。
DOI:
10.1109/embc.2014.6943732
复制
发表时间:
2014
期刊:
影响因子:
--
通讯作者:
Meng,Ellis
中科院分区:
文献类型:
--
作者:
Sheybani,Roya;Meng,Ellis
Wireless infusion rate control and programmability for an implantable, low power, electrochemical micropump is presented. Flow rate control was achieved through adjustment of the wiper position of a current potentiometer in the wireless receiver (0.6-3.2 mA output current with a resolution of 0.2 mA per step). An off-the-shelf Bluetooth module and Basic Stamp microcontroller kit was used to initiate amplitude-shift keying (ASK) modulation of the inductive power signal. Accurate flow control of two model regimens was achieved on benchtop. Wireless transmission (power transfer and control) was not affected by simulated tissue material placed between the transmitter and receiver.