In vitro artefact assessment of a new MR-compatible microwave antenna and a standard MR-compatible radiofrequency ablation electrode for tumour ablation
In vitro artefact assessment of a new MR-compatible microwave antenna and a standard MR-compatible radiofrequency ablation electrode for tumour ablation
复制标题
DOI:
10.1007/s00330-015-3891-0
复制
发表时间:
2016-03-01
影响因子:
5.9
通讯作者:
Clasen, Stephan
中科院分区:
文献类型:
--
作者:
Hoffmann, Ruediger;Rempp, Hansjoerg;Clasen, Stephan
To evaluate and compare artefact configuration and diameters in a magnetic resonance (MR)-compatible prototype microwave (MW) applicator and a standard MR-compatible radiofrequency (RF) applicator for MR-guided tumour ablation.Both applicators were tested in a phantom study at 1.5T with three sequences: T1-weighted three-dimensional volume interpolated breath-hold examination (VIBE), T1-weighted fast low angle shot (FLASH), T2-weighted turbo spin echo (TSE). Applicator orientation to main magnetic field (B-0) and slice orientation were varied. Needle tip location error (TLE) was assessed, and artefact diameters were calculated. Influence of imaging parameters on artefacts was assessed with analysis of variance (ANOVA) and post hoc testing.MW applicator: the shaft artefact diameter measured 2.3 +/- 0.8 mm. Tip artefact diameter and length measured 2.2 +/- 0.8 mm and 2.4 +/- 1.3 mm, respectively. A prominent oval artefact (diameter: 16.5 +/- 1.8 mm, length: 19.1 +/- 2.5 mm) appeared close to the tip. TLE: - .3 +/- 0.6 mm. RF applicator: shaft and tip diameter measured 8.9 +/- 4.7 mm and 9.0 +/- .0 mm, respectively. TLE: -0.1 +/- 0.8 mm. Minimal artefacts were measured with RF applicator orientation parallel to B-0 (P < 0.0001), whereas no such influence was found for MW applicator. For both applicators, significantly large artefacts were measured with T1 FLASH (P = 0.03).The MW applicator's artefact is satisfactory and seems useable for MR-guided ablation procedures.aEuro cent MW applicator's artefact appearance is independent of angulation to main magnetic field.aEuro cent MW applicator's prominent distal artefact may increase visibility under MR-guidance.aEuro cent RF and MW applicator's artefacts are precise concerning tip depiction.aEuro cent Largest artefact diameters are measured with T1-weighted fast low angle shot sequence.