SENSITIVITY, ACCURACY, AND SAFETY OF AN AUTOMATIC EXTERNAL DEFIBRILLATOR Report of a Field Evaluation

SENSITIVITY, ACCURACY, AND SAFETY OF AN AUTOMATIC EXTERNAL DEFIBRILLATOR Report of a Field Evaluation
复制标题

DOI:
10.1016/s0140-6736(84)92688-6
复制
发表时间:
1984-08
期刊:
The Lancet
影响因子:
--
通讯作者:
R. Cummins;L. Bergner;M. Eisenberg;John A. Murray
R. Cummins;L. Bergner;M. Eisenberg;John A. Murray
中科院分区:
其他
文献类型:
--
作者:
R. Cummins;L. Bergner;M. Eisenberg;John A. Murray

文献摘要

被引文献

相似文献

护理人员使用自动体外除颤器(AED)检测39例院外心脏骤停患者的心室颤动并给予反击。AED对16例心室颤动患者中的13例进行了识别并至少给予了1次反击(灵敏度为81%)。AED对所有21种非心室颤动节律(8种其他电节律,13种无骤停)均有正确反应,无反搏(100%特异性)。2例患者心律无法评估。该装置没有对患者或人员造成伤害。通过将每个15 s心室颤动段作为单独的挑战来分析AED的性能;该装置在29个这样的片段中有19个(66%)产生了反击。
An automatic external defibrillator (AED) was used by paramedics to detect ventricular fibrillation and deliver countershocks in 39 people with out-of-hospital cardiac arrests. The AED identified and delivered at least 1 countershock to 13 of the 16 people in ventricular fibrillation (81% sensitivity). The AED responded correctly to all 21 of the non-ventricular-fibrillation rhythms (8 other electrical rhythms, 13 asystole) with no countershocks (100% specificity). In 2 patients the rhythm could not be assessed. The device caused no injuries to patients or personnel. The performance of the AED was also analysed by considering each 15 s segment of ventricular fibrillation as a separate challenge; the device delivered a countershock in 19 of 29 such segments (66%).