SHAPE ESTIMATION OF BOWTIE FILTERS BASED ON THE LUMINESCENCE FROM POLYETHYLENE TEREPHTHALATE RESIN BY X-RAY IRRADIATION
SHAPE ESTIMATION OF BOWTIE FILTERS BASED ON THE LUMINESCENCE FROM POLYETHYLENE TEREPHTHALATE RESIN BY X-RAY IRRADIATION
复制标题
基于聚对苯二甲酸乙二醇酯树脂 X 射线辐射发光的蝶形滤波器形状估计
DOI:
10.1093/rpd/ncz031
复制
发表时间:
2019
影响因子:
1
通讯作者:
N.
中科院分区:
文献类型:
--
作者:
Yamauchi-Kawaura;C.;Fujii;K.;Yamauchi;M.;Yamamoto;S.;Kozuka;M.;Ohzawa;N.;Suga;N. and Ito;N.
In this study, we devised a novel method estimating the bowtie filter shapes by imaging luminescence from a polyethylene terephthalate (PET) resin with X-ray irradiation in a computed tomography (CT) scanner. The luminescence distribution of the PET resin corresponding to the thickness of bowtie filter was imaged using a charge-coupled device camera. On the assumption that the material of bowtie filter is aluminium (Al), the shape of bowtie filters was estimated from the correlation between Al attenuation curves and the angular-dependent luminance attenuation profiles according to the thickness of bowtie filters. Dose simulations based on the estimated bowtie filter shapes were performed using head and body PMMA phantoms with 16 and 32 cm in diameter. The simulated values of head and body weighted CT dose index (CTDIw) based on bowtie filter shape by the luminescence imaging method agreed within ~9% with the measured values by a dosemeter.