ADAPTIVE MULTIVARIABLE DRUG DELIVERY - CONTROL OF ARTERIAL-PRESSURE AND CARDIAC-OUTPUT IN ANESTHETIZED DOGS
ADAPTIVE MULTIVARIABLE DRUG DELIVERY - CONTROL OF ARTERIAL-PRESSURE AND CARDIAC-OUTPUT IN ANESTHETIZED DOGS
复制标题
DOI:
10.1109/tbme.1987.326093
复制
发表时间:
1987-08-01
影响因子:
4.6
通讯作者:
CHIZECK, HJ
中科院分区:
文献类型:
--
作者:
VOSS, GI;KATONA, PG;CHIZECK, HJ
An adaptive algorithm (CAMAC, control advance moving average controller) to control multiinput/multioutput physiological systems has been implemented and tested. The algorithm is a self-tuning controller that determines the input on the basis of the expected difference between the output and desired output at a time interval equal to or greater than the system dead time. The algorithm was used to simultaneously control mean arterial pressure and cardiac output in anesthetized dogs by the simultaneous computer-controlled infusion of sodium nitroprusside and dobutamine. The results demonstrate the feasibility of using CAMAC for multivariable drug delivery, but they indicate the need for further work before clinical applications are attempted.