The Spectrum of Th-Ar Hollow Cathode Lamps in the 691-5804 nm region: Establishing Wavelength Standards for the Calibration of Infrared Spectrographs
The Spectrum of Th-Ar Hollow Cathode Lamps in the 691-5804 nm region: Establishing Wavelength Standards for the Calibration of Infrared Spectrographs
复制标题
Th-Ar 空心阴极灯在 691-5804 nm 区域的光谱:建立红外光谱仪校准的波长标准
DOI:
10.1086/590111
复制
发表时间:
2008
期刊:
影响因子:
--
通讯作者:
C. Sansonetti
中科院分区:
文献类型:
--
作者:
F. Kerber;Gillian Nave;C. Sansonetti
We report new observations of the infrared (IR) spectrum of low-current Th-Ar hollow cathode lamps with the 2 m Fourier transform spectrometer (FTS) at the National Institute of Standards and Technology. These observations establish more than 2400 lines that are suitable for use as wavelength standards in the range 900-4500 nm. The line list is used as input for a physical instrument model that provides the wavelength calibration for the Cryogenic High-Resolution IR Echelle Spectrometer (CRIRES), the European Southern Observatory’s new high-resolution (R ∼ 100,000) IR spectrograph at the Very Large Telescope. We have also observed the variation of the spectrum of Th-Ar lamps as a function of operating current. The results allow us to optimize the spectral output in terms of relative intensity and line density for operation on the telescope. Our results should be generally useful for wavelength calibration in near-IR astronomy, providing a high density of sharp, well-characterized emission lines with the ease and efficiency of operation of a commercial discharge lamp.