Validation of the CAS neonatal NIRS system by monitoring vv-ECMO patients: preliminary results.
Validation of the CAS neonatal NIRS system by monitoring vv-ECMO patients: preliminary results.
复制标题
通过监测 vv-ECMO 患者验证 CAS 新生儿 NIRS 系统:初步结果。
DOI:
10.1007/0-387-26206-7_27
复制
发表时间:
2005
影响因子:
--
通讯作者:
Short,BillieL
中科院分区:
文献类型:
--
作者:
Benni,PaulB;Chen,Bo;Dykes,FrancineD;Wagoner,ScottF;Heard,Micheal;Tanner,AprilJ;Young,TeresaL;Rais-Bahrami,Khodayar;Rivera,Oswaldo;Short,BillieL
The CAS neonatal NIRS system determines absolute regional brain tissue oxygen saturation (SnO2) and brain true venous oxygen saturation (SnvO2) non-invasively. Since NIRS-interrogated tissue contains both arterial and venous blood from arterioles, venules, and capillaries, SnO2is a mixed oxygen saturation parameter, having values between arterial oxygen saturation (SaO2) and cerebral venous oxygen saturation (SVO2). To determine a reference for SnO2, the relative contribution of SvO2to SaO2drawn from a brain venous site vs. systemic SaO2is approximately 70:30 (SvO2:SaO2). If the relationship of the relative average contribution of SvO2and SaO2is known and does not change to a large degree, then NIRS true venous oxygen saturation, SnvO2, can be determined non-invasively using SnO2along with SaO2from a pulse oximeter.