Determination of melamine in powdered milk by molecularly imprinted stir bar sorptive extraction coupled with HPLC.
Determination of melamine in powdered milk by molecularly imprinted stir bar sorptive extraction coupled with HPLC.
复制标题
DOI:
10.1016/j.jcis.2015.05.008
复制
发表时间:
2015-09
影响因子:
9.9
通讯作者:
Ling Zhu;Guanhong Xu;Fangdi Wei;Jing Yang;Qin Hu
中科院分区:
文献类型:
--
作者:
Ling Zhu;Guanhong Xu;Fangdi Wei;Jing Yang;Qin Hu
A novel molecularly imprinted stir bar (MIP-SB) was developed with melamine (MA) as the template molecule in this study. The sorptive capacity of MIP-SB was nearly three times over that of non-imprinted stir bar (NIP-SB). The MIP-SB presented much better selectivity than NIP-SB when used to absorb MA and its analogues. An analytical method to determine MA in powdered milk was established by combining MIP-SB sorptive extraction with HPLC. The liner range was 0.0631–12.6 ng/mL with good correlation coefficient of 0.9983, and the limit of detection was 0.0127 ng/mL based on three times ratio of signal to noise. This method was successfully applied to the determination of MA in powdered milk with satisfactory results. The method built is simple and suitable for the determination of MA in milk products which is of great significance for quality control of milk products.