A comparison of in-air and in-water calibration of a dosimetry system used for radiation dose assessment in cancer therapy
A comparison of in-air and in-water calibration of a dosimetry system used for radiation dose assessment in cancer therapy
复制标题
DOI:
10.2298/ntrp1001051a
复制
发表时间:
2010
影响因子:
1.2
通讯作者:
W. Arshed;Kh. Mahmood;I. Qazi;A. Ullah;P. Akhter;Salman Ahmad;Zeeshan Jamil
中科院分区:
文献类型:
--
作者:
W. Arshed;Kh. Mahmood;I. Qazi;A. Ullah;P. Akhter;Salman Ahmad;Zeeshan Jamil
An accurate calibration of the therapy level radiation dosimetry system has a pivotal role in the accuracy of dose delivery to cancer patients. The two methods used for obtaining a tissue equivalent calibration of the system: air kerma calibration and its conversion to a tissue equivalent value (absorbed dose to water) and direct calibration of the system in a water phantom, have been compared for identical irradiation geometry. It was found that the deviation between the two methods remained within a range of 0% to ±1.7% for the PTW UNIDOS dosimetry system. This means that although the recommended method is in-water calibration, under exceptional circumstances, in-air calibration may be used as well.