Fast kinetic measurements and on-line dilution by flow injection cytometry.
Fast kinetic measurements and on-line dilution by flow injection cytometry.
复制标题
通过流式注射细胞术进行快速动力学测量和在线稀释。
DOI:
10.1002/cyto.990160406
复制
发表时间:
1994
期刊:
影响因子:
--
通讯作者:
Christian,GD
中科院分区:
文献类型:
--
作者:
Lindberg,W;Scampavia,LD;Ruzicka,J;Christian,GD
An improved flow injection cytometry (FIC) system suitable for fast cellular kinetic measurements and on‐line dilution is described. The instrument allows for measurements from 1.2 s (± 0.05 s) after the initiation of mixing, up to any time thereafter. Crucial factors in determining fast kinetic measurements, such as the displacement volume for sample introduction, rate of transport to cytometric interrogation point, and the mixing speed are evaluated and discussed. By varying the volume aspirated, this instrument can facilitate the dilution of cells and/or reagent over a range of one order of magnitude. The fast kinetic and on‐line dilution capabilities were demonstrated by on‐line staining of DNA in trout erythrocytes with 4′, 6‐diamidino‐2‐phenylindole (DAPI). The utility of the instrument for measurement of enzyme kinetics was illustrated using human lymphocytes, measuring the glutathione S‐transferase (GST) catalyzed reaction between monochlorobimane (MCB) and glutathione (GSH). © 1994 Wiley‐Liss, Inc.