Positron emission mammographic instrument: initial results.
Positron emission mammographic instrument: initial results.
复制标题
正电子发射乳腺X线摄影仪:初步结果。
作者:
K. Murthy;M. Aznar;A. Bergman;Christopher J. Thompson;J. Robar;Robert Lisbona;Antoine Loutfi;Jean H. Gagnon
Performance characteristics of a positron emission mammographic (PEM) instrument were studied. This dedicated metabolic breast imaging system has spatial resolution of 2.8-mm full width at half maximum (FWHM), coincidence resolving time of 12-nsec FWHM, and absolute efficiency of 3%. Hot spots with diameter of 16 mm in a phantom with signal-to-background activity ratio of 6:1 were distinguishable with a scanning time of 5 minutes.