A critical study on circular dichroism measurement in longer side of visible region.
A critical study on circular dichroism measurement in longer side of visible region.
复制标题
可见光区长边圆二色性测量的批判性研究。
DOI:
--
复制
发表时间:
1981
期刊:
影响因子:
--
通讯作者:
M. Hatano
中科院分区:
文献类型:
--
作者:
T. Konno;H. Meguro;T. Murakami;M. Hatano
A test sample method revealed large instrumental deviations of molar ellipticities, [Θ], up to ±35% in circular dichroism(CD) in the 500-800 nm region, even after (+)-10-camphorsulfonic acid(CSA) calibration at 290 nm. Use of the nickel (II)tartrate solution and the tentative [Θ] values, −101.2 at 718 nm and −110.7 at 777 nm(degree dm3 · mol−1cm−1), were recommended to calibrate the wavelengths and magnitudes of the CD extrema in this region.