Liver acquisition with volume acceleration flex on 70-cm wide-bore and 60-cm conventional-bore 3.0-T MRI

Liver acquisition with volume acceleration flex on 70-cm wide-bore and 60-cm conventional-bore 3.0-T MRI
复制标题

DOI:
10.1007/s12194-015-0344-z
复制
发表时间:
2016-07-01
影响因子:
1.6
通讯作者:
Hashido, Takashi
Hashido, Takashi
中科院分区:
其他
文献类型:
--
作者:
Saito, Shigeyoshi;Tanaka, Keiko;Hashido, Takashi

文献摘要

被引文献

相似文献

本研究旨在比较在60 cm常规孔径和70 cm宽孔径3.0 T磁共振成像(MRI)上使用容积加速Flex(LAVA-Flex)和LAVA进行肝脏采集的脂肪抑制均匀性和图像质量。LAVA-Flex和LAVA的脂肪抑制均匀性被评估为在肝顶、门和肾门水平抑制浅表脂肪的效率。使用以下公式计算百分比标准偏差(%SD):%SD(%)= 100 x感兴趣区域(ROI)的SD/ROI中信号强度(SI)的平均值。计算信噪比(SNR)和对比度(CR)。在LAVA序列中,宽孔3.0-T MRI所有层面的%SD均显著高于常规孔3.0-T MRI(P
This study aimed to compare the uniformity of fat suppression and image quality between liver acquisition with volume acceleration flex (LAVA-Flex) and LAVA on 60-cm conventional-bore and 70-cm wide-bore 3.0-T magnetic resonance imaging (MRI). The uniformity of fat suppression by LAVA-Flex and LAVA was assessed as the efficiency of suppression of superficial fat at the levels of the liver dome, porta, and renal hilum. Percentage standard deviation (%SD) was calculated using the following equation: %SD (%) = 100 x SD of the regions of interest (ROIs)/mean value of the signal intensity (SI) in the ROIs. Signal-to-noise ratio (SNR) and contrast ratio (CR) were calculated. In the LAVA sequence, the %SD in all slices on wide-bore 3.0-T MRI was significantly higher than that on conventional-bore 3.0-T MRI (P