Magnetic Dynabeads® detection by sensitive element based on giant magnetoimpedance
Magnetic Dynabeads® detection by sensitive element based on giant magnetoimpedance
复制标题
DOI:
10.1016/j.bios.2004.04.027
复制
发表时间:
2005-02-15
影响因子:
12.6
通讯作者:
Levit, V
中科院分区:
文献类型:
--
作者:
Kurlyandskaya, G;Levit, V
A GMI-biosensor prototype was designed and fabricated using an amorphous ribbon. As a first step, a GMI response was measured with a model liquid media: suspension containing magnetisable Dynabeads(R) M-450. These beads can be used as biomolecular labels when coated with a specific antibody. The GMI change caused by their presence was measured with a biosensor prototype designed with a measuring cell containing an amorphous CoFeMoSiB-ribbon element. GMI was measured at a range of current frequencies from 0.3 to 10 MHz and at intensities of either I-rms = 3 or 4.5 mA. Commercial Dynabeads(R) M-450 were supplied as a suspension in a phosphate buffered saline. A maximum difference of 25% in the GMI ratio measured with and without Dynabeads(R) was obtained at a frequency of 3.5 MHz and 1 = 3 mA. Some potential applications for GMI-biosensor and further directions are proposed. (C) 2004 Elsevier B.V. All rights reserved.