The use of high-performance liquid chromatography for quantitation of plasma amino acids in man.
The use of high-performance liquid chromatography for quantitation of plasma amino acids in man.
复制标题
使用高效液相色谱法定量人体血浆氨基酸。
DOI:
10.1016/0003-2697(82)90138-5
复制
发表时间:
1982
影响因子:
2.9
通讯作者:
Isenberg,JI
中科院分区:
文献类型:
--
作者:
Hogan,DL;Kraemer,KL;Isenberg,JI
Precolumn derivatization of primary amino acids with o-phthaldialdehyde forms fluorescent derivatives that can be separated by reverse-phase high-performance liquid chromatography (HPLC). Human plasma deproteinized and then derivatized by this method produced individual amino acid peaks in less than 35 min. Quantitative analysis of micro amounts of plasma-free amino acids gave highly reproducible results (mean coefficient of variation 2.9 ± 0.5%; r = 0.988 ± 0.005; p < 0.001). There was good agreement between measurement of plasma amino acids with HPLC and Beckman 119 and 120B amino acid analyzers; coefficient of variation 13.4 ± 4.3% and 19.0 ± 3.6%, respectively; r = 0.936 and 0.711, respectively (p < 0.05). There was a significant increase in plasma amino acids after a protein meal. HPLC is a rapid, sensitive, and precise method of measuring plasma amino acids in man.