Agreement between Time-Domain and Spectral-Domain Optical Coherence Tomography in the Assessment of Macular Thickness in Patients with Idiopathic Macular Telangiectasia Type 2
Agreement between Time-Domain and Spectral-Domain Optical Coherence Tomography in the Assessment of Macular Thickness in Patients with Idiopathic Macular Telangiectasia Type 2
复制标题
DOI:
10.1159/000353455
复制
发表时间:
2013-01-01
期刊:
影响因子:
2.6
通讯作者:
Peto, Tunde
中科院分区:
文献类型:
--
作者:
Mathew, Raeba;Sivaprasad, Sobha;Peto, Tunde
Purpose: To estimate the conversion factors to transpose macular thickness measurements on time-domain (TD) to various spectral-domain (SD) optical coherence tomography (OCT) machines in patients with macular telangiectasia type 2a (MacTel). Procedures: Macular scans on TD- and SD-OCT were performed on patients at the same visit. The retinal thickness values in various ETDRS subfields and macular volume were compared between different OCT machines. Results: The macular thickness and volume were significantly greater (p < 0.0001, r = 0.678-0.822) on SD-OCT. The mean differences in macular thickness between TD Stratus and SD-OCT by Spectralis, Cirrus and Topcon were 62, 41 and 20 mu m, respectively. The conversion factor of macular thickness from TD-OCT to Spectralis, Cirrus and Topcon were +65, +39 and +25 mu m, respectively. Conclusion and Message: The estimates of conversion of macular thickness from TD- to SD-OCT using simple mean differences between machines and those by linear regression were similar suggesting that the former may be used for the longitudinal follow-up of MacTel patients. (C) 2013 S. Karger AG, Basel