Reliability of musculoskeletal ultrasound imaging to measure supraspinatus tendon thickness in healthy subjects.

Reliability of musculoskeletal ultrasound imaging to measure supraspinatus tendon thickness in healthy subjects.
复制标题

肌肉骨骼超声成像的可靠性,以测量健康受试者中肌腱厚度。

DOI:
10.1589/jpts.29.1394
复制
发表时间:
2017-08
影响因子:
--
通讯作者:
Tanveer F
Tanveer F
中科院分区:
其他
文献类型:
--
作者:
Ahmad A;Bandpei MAM;Gilani SA;Munawar A;Ahmed I;Tanveer F

文献摘要

被引文献

相似文献

[目的] 本研究的目的是评估肌肉骨骼超声 (MSKUS) 作为测量健康受试者冈上肌腱厚度的工具的评估者内和评估者间的可靠性。 [对象和方法] 20名健康受试者以相似的坐姿进行评估。两位经验丰富的肌肉骨骼超声专家测量了冈上肌腱的厚度。两名评估者在一天内对肌腱进行两次测量,每次间隔一小时,以获得日内可靠性,并在一周后重新测量日间可靠性。使用组内相关系数(ICC)和测量标准误差(SEM)进行数据分析。 [结果] 结果反映了出色的日内评估者内和评估者间可靠性,平均值±标准差 (SD)=5.14±0.83 和 5.17±0.78,ICC (2,1)=0.89–0.93 和 ICC (2,2)=0.98 (0.95−0.99),SEM=0.18–0.18 和 0.17。评估者内和评估者间信度均值 ± SD=5.14 ± 0.83 和 5.17 ± 0.78,ICC(2,1)=0.88–0.90 和 ICC(2,2)=0.95 (0.88−0.98),SEM=0.17–0.38 和 0.17 【结论】当前研究结果表明 MSKUS 是可靠的用于测量健康受试者冈上肌腱厚度的工具。
[Purpose] The purpose of this study was to assess intra-rater and inter-rater reliability of musculoskeletal ultrasound (MSKUS) as a tool to measure supraspinatus tendon thickness in healthy subjects. [Subjects and Methods] Twenty healthy subjects were assessed in similar sitting position. Two experienced musculoskeletal ultrasound specialists measured supraspinatus tendon thickness. Tendon was measured twice in one day by each of two raters with an interval of one hour for within day reliability and was re-measured following one week for between days reliability. Data analysis was performed using the intraclass correlation coefficient (ICC) and standard error of measurement (SEM). [Results] Results reflected excellent within day intra-rater and inter-rater reliability with Mean ± standard deviation (SD)=5.14 ± 0.83 and 5.17 ± 0.78, ICC (2,1)=0.89–0.93 and ICC (2,2)=0.98 (0.95−0.99) and SEM=0.18–0.18 and 0.17 while between days intra-rater rater and inter-rater reliability with Mean ± SD=5.14 ± 0.83 and 5.17 ± 0.78, ICC(2,1)=0.88–0.90 and ICC(2,2)=0.95 (0.88−0.98), SEM=0.17–0.38 and 0.17 [Conclusion] The results of the current study indicate that MSKUS is a reliable tool for measuring supraspinatus tendon thickness in healthy subjects.