Quantitative analysis of amylopectin unit chains by means of high-performance anion-exchange chromatography with pulsed amperometric detection
Quantitative analysis of amylopectin unit chains by means of high-performance anion-exchange chromatography with pulsed amperometric detection
复制标题
DOI:
10.1016/s0021-9673(97)01151-5
复制
发表时间:
1998-03-27
影响因子:
4.1
通讯作者:
Åman, P
中科院分区:
文献类型:
--
作者:
Koch, K;Andersson, R;Åman, P
High-performance anion-exchange chromatography with pulsed amperometric detection (HPAEC-PAD) is a widely used technique to study the chain length distribution of amylopectin. The chromatograms, however, do not directly reflect this distribution, since the PAD response changes with the degree of polymerization. In this study, waxy maize starch was debranched and fractionated on a Bio-Gel P-6 column and the response factors for maltosaccharides with DP 3-65 were determined. The detector response per mu g glucan chains was shown to decrease considerably for DP 3-7 while the curve leveled out for DP larger than 15. (C) 1998 Elsevier Science B.V.