The accuracy of the FreeStyle navigator continuous glucose monitoring system in children with type 1 diabetes
The accuracy of the FreeStyle navigator continuous glucose monitoring system in children with type 1 diabetes
复制标题
DOI:
10.2337/dc06-1407
复制
发表时间:
2007-01-01
期刊:
影响因子:
16.2
通讯作者:
Weinzimer, Stuart A.
中科院分区:
文献类型:
--
作者:
Wilson, Darrell M.;Beck, Roy W.;Weinzimer, Stuart A.
OBJECTIVE - To evaluate the accuracy and precision of the FreeStyle Navigator continuous glucose monitoring system in children with type I diabetes.RESEARCH DESIGN AND METHODS - in 30 children with type I diabetes (mean age 11.2 +/- 4.1 years), the Navigator glucose values were compared with reference serum glucose values of blood samples obtained in an inpatient clinical research center and measured in a central laboratory using a hexokinase enzymatic method and in an outpatient setting with a FreeStyle meter. Median absolute difference (AD) and median relative absolute difference (RAD) were computed for sensor-reference and sensor-sensor pairs.RESULTS - The median AD and RAD were 17 mg/dl and 12%, respectively, for 1,811 inpatient sensor-reference pairs and 20 mg/dl and 14%, respectively, for 8,639 outpatient pairs. The median RAD between two simultaneous Navigator measurements (n = 1,971) was 13%, Ninety-one percent of sensors in the inpatient setting and 81% of sensors in the outpatient setting had a median RAD