Quantum efficiency of the CCD camera (XIS) for the ASTRO-E mission
Quantum efficiency of the CCD camera (XIS) for the ASTRO-E mission
复制标题
DOI:
10.1016/s0168-9002(99)00600-2
复制
发表时间:
1999-10-21
影响因子:
1.4
通讯作者:
Kissel, S
中科院分区:
文献类型:
--
作者:
Katayama, H;Shouho, M;Kissel, S
We measured the optical and the X-ray transmission of the optical blocking filters for the X-ray Imaging Spectrometers (XISs) which are the X-ray CCD cameras of the ASTRO-E satellite. We conclude that the oxidation of the aluminum reduces the optical transmission down to similar to 60-70% of the theoretical value of the aluminum. We achieved optical transmission below 5 x 10(-5) in the range from 4000 to 9500 Angstrom by using aluminum thickness of 1200 Angstrom, while the theoretical calculation requires 800 Angstrom. The measurement of absolute quantum efficiency of XIS is also performed at several particular energies. We confirmed 20% quantum efficiency at 0.5 keV for the XIS engineering model (XIS EM). (C) 1999 Elsevier Science B.V. All rights reserved.