Blood flow occlusion via ultrasound image-guided high-intensity focused ultrasound and its effect on tissue perfusion
Blood flow occlusion via ultrasound image-guided high-intensity focused ultrasound and its effect on tissue perfusion
复制标题
DOI:
10.1016/j.ultrasmedbio.2006.08.016
复制
发表时间:
2007-03-01
影响因子:
2.9
通讯作者:
Okai, Takashi
中科院分区:
文献类型:
--
作者:
Ichihara, Mitsuyoshi;Sasaki, Kazuaki;Okai, Takashi
This study investigated the induction of tissue necrosis by arterial blood flow occlusion using ultrasound image-guided high-intensity focused ultrasound (HIFU). We constructed a prototype HIFU transducer in combination with an imaging probe that provided color Doppler imaging and ultrasound contrast imaging. The HIFU beam was aimed into a branch of the renal artery in vivo. The renal artery branches of eight rabbits were occluded by HIFU at an intensity of 4 kW/cm(2) (from 2 to 10 times of each sonication for 5 s). When the HIFU exposure was successful, complete cessation of blood flow was observed by color Doppler imaging with success rate of 100% (8/8). Furthermore, lack of perfusion was observed in the renal cortex with a contrast-enhanced image. Postmortem histologic evaluation showed a wedge-shaped area of infarction in six of seven cases, corresponding to the lack of the contrast medium in the ultrasound image. These results demonstrated that ultrasound image-guided HIFU can be used to induce arterial occlusion, thus producing infarction and necrosis of the perfused tissue. (E-mail: m_itihara@hotmail.com) (c) 2007 World Federation for Ultrasound in Medicine & Biology.