A measurement of the scintillation decay time constant in liquid xenon with the XMASS-I detector
A measurement of the scintillation decay time constant in liquid xenon with the XMASS-I detector
复制标题
使用 XMASS-I 探测器测量液氙中的闪烁衰减时间常数
DOI:
10.1088/1742-6596/1468/1/012235
复制
发表时间:
2020
期刊:
影响因子:
--
通讯作者:
Ichimura Koichi
中科院分区:
文献类型:
--
作者:
青木貴稔;鳥井寿夫;Sahoo Bijaya Kumar;Das Bhanu Pratap;原田健一;早水友洋;坂本幸祐;川村広 和;井上壮志;内山愛子;伊藤沙希;吉岡里紗;田中香津生;伊藤正俊;畠山温;長濱弘季;小澤直也; 酒見泰寛;Ichimura Koichi
The scintillation decay time constant in liquid xenon was measured using the XMASS-I detector and its calibration setup. The scintillation decay time constants of both electron recoil and nuclear recoil were evaluated by comparing the observed photon detection times with Monte Carlo simulations. Two exponential components, corresponding to the singlet and triplet state of the xenon excited dimer, are needed to reproduce the observed photon detection times. In this paper, we report the measurement of the electron recoil and the nuclear recoil scintillation decay time constants with the XMASS-I detector.