INTERCOMPARISON OF PERMITTIVITY MEASUREMENTS USING THE TRANSMISSION/REFLECTION METHOD IN 7-MM COAXIAL TRANSMISSION-LINES
INTERCOMPARISON OF PERMITTIVITY MEASUREMENTS USING THE TRANSMISSION/REFLECTION METHOD IN 7-MM COAXIAL TRANSMISSION-LINES
复制标题
DOI:
10.1109/22.330120
复制
发表时间:
1994-11-01
影响因子:
4.3
通讯作者:
JANEZIC, MD
中科院分区:
文献类型:
--
作者:
VANZURA, EJ;BAKERJARVIS, JR;JANEZIC, MD
Broadband permittivity measurements made by eleven organizations using the transmission/reflection (T/R) method are compared to high-accuracy cavity resonator results. T/R accuracy is less than 10% for epsilon'(r) < 15, and the smallest measureable loss factor is epsilon''(r) approximate to 0.05. Uncertainty caused by the air gaps between the specimen and the inner and outer conductors is the largest contributor to the overall uncertainty. Compared to other dimensional measurement methods, physical measurement of specimen bore and outer diameters yield the most accurate gap corrections.