Demonstration of dose and scatter reductions for interior computed tomography.
Demonstration of dose and scatter reductions for interior computed tomography.
复制标题
内部计算机断层扫描的剂量和散射减少演示。
DOI:
10.1097/rct.0b013e31819f61e2
复制
发表时间:
2009
影响因子:
1.3
通讯作者:
Wang,Ge
中科院分区:
文献类型:
--
作者:
Bharkhada,Deepak;Yu,Hengyong;Dixon,Robert;Wei,Yuchuan;Carr,JJeffrey;Bourland,JDaniel;Best,Ryan;Hogan,Richard;Wang,Ge
With continuing developments in computed tomography (CT) technology and its increasing use of CT imaging, the ionizing radiation dose from CT is becoming a major public concern particularly for high-dose applications such as cardiac imaging. We recently proposed a novel interior tomography approach for x-ray dose reduction that is very different from all the previously proposed methods. Our method only uses the projection data for the rays passing through the desired region of interest. This method not only reduces x-ray dose but scatter as well. In this paper, we quantify the reduction in the amount of x-ray dose and scattered radiation that could be achieved using this method. Results indicate that interior tomography may reduce the x-ray dose by 18% to 58% and scatter to the detectors by 19% to 59% as the FOV is reduced from 50 to 8.6 cm.