A high resolution GC-analysis method for determination of C2-C10 hydrocarbons in air samples
A high resolution GC-analysis method for determination of C2-C10 hydrocarbons in air samples
复制标题
测定空气样品中 C2-C10 碳氢化合物的高分辨率 GC 分析方法
DOI:
10.1002/jhrc.1240091007
复制
发表时间:
1986
期刊:
影响因子:
--
通讯作者:
W. Seiler
中科院分区:
文献类型:
--
作者:
P. Matuška;M. Koval;W. Seiler
An analytical method for the determination of non-methane hydrocarbons (NMHC) in air was developed using simultaneously pressure and temperature programmed gas chromatography (GC) on a capillary column. The separation of C2-C10 hydrocarbons was performed in a single GC-run using a fused silica capillary column (FSCC) with a chemically bonded non-polar stationary phase (CP-Sil-5-CB). A combination of on-column cryofocusing and GC reinjection provided baseline separation of C2 hydrocarbons at subambient initial temperature (−40°C). The cryofocusing was performed on the column head immersed in liquid nitrogen. Quantitative retention of all C2-C10 hydrocarbons was achieved only when a temperature gradient is created along the cooled column section. The focused components are eluted from the column head without supplementary heating by removing the coolant. To eliminate matrix effects, the enrichment procedure was designed to remove CO2 and H2O. The detection limit was found to be 2.0 × 10−12 g propane.