Assessment of chest CT at CTDIvol less than 1mGy with iterative reconstruction techniques

Assessment of chest CT at CTDIvol less than 1mGy with iterative reconstruction techniques
复制标题

DOI:
10.1259/bjr.20160625
复制
发表时间:
2017-01-01
影响因子:
2.6
通讯作者:
Kalra, Mannudeep K.
Kalra, Mannudeep K.
中科院分区:
医学3区
文献类型:
--
作者:
Padole, Atul;Digumarthy, Subba;Kalra, Mannudeep K.

文献摘要

被引文献

相似文献

目的:评价基于图像的迭代重建(SafeCT;MedicVision(R),Tirat Carmel,以色列)、自适应统计迭代重建(ASIR;GE Healthcare,Waukeha,WI)和基于模型的迭代重建(MBIR;GE Healthcare,Waukeha,WI)技术在CT剂量指数体积(CTDIval)下重建胸部CT的图像质量
Objective: To assess the image quality of chest CT reconstructed with image-based iterative reconstruction (SafeCT; MedicVision (R), Tirat Carmel, Israel), adaptive statistical iterative reconstruction (ASIR; GE Healthcare, Waukesha, WI) and model-based iterative reconstruction (MBIR; GE Healthcare, Waukesha, WI) techniques at CT dose index volume (CTDIvol)