Experimental demonstration of noncontact pulse wave velocity monitoring using multiple Doppler radar sensors
Experimental demonstration of noncontact pulse wave velocity monitoring using multiple Doppler radar sensors
复制标题
使用多个多普勒雷达传感器进行非接触式脉搏波速度监测的实验演示
DOI:
--
复制
发表时间:
2010
期刊:
影响因子:
--
通讯作者:
D. Lie
中科院分区:
文献类型:
--
作者:
Li Lu;Changzhi Li;D. Lie
In this paper, two Doppler radars are used to monitor the pulse movements at the heart and the calf in order to measure the pulse wave velocity (PWV) wirelessly. Both simulation and experiment have been performed to demonstrate the feasibility of the proposed noncontact PWV monitoring. A three-stage calibration procedure, including DC offset calibration, circuit delay calibration and antenna radiation pattern calibration, has been developed for reliable long-term PWV monitoring. The measurement results have been verified by wired contact measurement with pulse transducers.