Rational-Gain-Compensation for Attenuation in Ultrasonic Cardiac Imaging
Rational-Gain-Compensation for Attenuation in Ultrasonic Cardiac Imaging
复制标题
超声心脏成像衰减的合理增益补偿
DOI:
10.1109/ultsym.1981.197695
复制
发表时间:
1981
影响因子:
6.7
通讯作者:
D. Skorton
中科院分区:
文献类型:
--
作者:
H. Melton;D. Skorton
The first step toward direct, reliable detection of myocardial infarction in vivo is a rational accounting for attenuation present between the transducer and sites of interest within the heart. One method was tested which compensated for attenuation on a line-by-line basis in real-time, two-dimensional ultrasonic scans. This method depends on the fact that ultrasonic scatter from blood is much smaller (absent) than scatter from myocardium, whether normal or abnormal. When back-scatter was "absent", receiver gain was increased at a rate suitable to compensate for attenuation by blood; when backscatter was "present", receiver gain was changed at a rate needed to compensate for attenuation by myocardium. This method provides a rational basis for comparison of back-scatter levels between different regions of myocardium.