Use of Stepped Porous Titanium Metaphyseal Sleeves for Tibial Defects in Revision Total Knee Arthroplasty: Short Term Results
Use of Stepped Porous Titanium Metaphyseal Sleeves for Tibial Defects in Revision Total Knee Arthroplasty: Short Term Results
复制标题
在翻修全膝关节置换术中使用阶梯式多孔钛骺套治疗胫骨缺损: 短期结果
DOI:
10.1016/j.arth.2013.12.026
复制
发表时间:
2014-06-01
影响因子:
3.5
通讯作者:
Gorab, Robert S.
中科院分区:
文献类型:
--
作者:
Barnett, Steven L.;Mayer, Ryan R.;Gorab, Robert S.
Stepped porous titanium metaphyseal sleeves may provide an option for enhanced fixation in managing challenging tibial defects in revision total knee arthroplasty (TKA). We retrospectively reviewed data on 51 patients who underwent revision TKA utilizing a metaphyseal sleeve for Anderson Orthopaedic Research Institute (AORI) Type II and III tibial defects between June 2007 and July 2011. Of these 51 patients, 36 patients had complete clinical and radiographic data. At final follow-up (mean: 38 months) significant improvements in knee range of motion and Knee Society scores were observed postoperatively (P < 0.001). Four revision procedures were necessary, but none for aseptic implant fixation failure. Radiographic review at final follow-up revealed stable, osteointegrated components without component migration or clinically significant osteolysis. Metaphyseal sleeve use in the management of moderate to severe tibial defects in revision TKA resulted in satisfactory clinical outcomes and is a versatile option for achieving stable fixation. (C) 2014 Elsevier Inc. All rights reserved.