Comparison of the Automated Random Access Immunoassay Analysers, ACS-180 (Ciba Corning) and AIA-1200 (Tosoh)
Comparison of the Automated Random Access Immunoassay Analysers, ACS-180 (Ciba Corning) and AIA-1200 (Tosoh)
复制标题
自动随机存取免疫分析仪 ACS-180 (Ciba Corning) 和 AIA-1200 (Tosoh) 的比较
DOI:
10.1515/cclm.1993.31.10.701
复制
发表时间:
1993
期刊:
影响因子:
--
通讯作者:
P. Janson
中科院分区:
文献类型:
--
作者:
G. Costongs;P. Janson
Two random access immunoassay analysers, the ACS-180 (Ciba Corning) and the AIA-1200 (Tosoh Corp.) were compared with respect to performance and user-friendliness. Precision studies revealed an almost equal intra-assay variation coefficient, but day-to-day reproducibility was better on the AIA-1200. Recovery of dilution series of tumour markers was too high with the ACS-180. Both systems were free from carry-over effects, on account of the extra wash step in the ACS-180 (for thyrotropin, carcinoembryonic antigen and human chorionic gonadotropin) and the separate pipetting tip in the AIA-1200. Reagent stability is better in the AIA-1200 than in the ACS-180. The AIA-1200 is more user-friendly and produces a higher throughput than the ACS-180 in daily routine operation (92 results/hour vs. 78 results/hour).