Improvement in glycemic excursions with a transcutaneous, real-time continuous glucose sensor
Improvement in glycemic excursions with a transcutaneous, real-time continuous glucose sensor
复制标题
DOI:
10.2337/diacare.29.1.44
复制
发表时间:
2006-01-01
期刊:
影响因子:
16.2
通讯作者:
Jovanovic, L
中科院分区:
文献类型:
--
作者:
Garg, S;Zisser, H;Jovanovic, L
OBJECTIVE - Hypoglycemia and wide glucose excursions continue to be major obstacles to achieving target HbA(1c) values and the associated reductions in long-term complications (and economic costs) in people with insulin-treated diabetes. In this study we evaluated the accuracy, safety, and clinical effectiveness of a continuous glucose-sensing device.RESEARCH DESIGN AND METHODS - A total of 91 insulin-requiring patients with type 1 (n = 75) and type 2 (n = 16) diabetes were enrolled in this multicenter randomized study. Subjects wore a transcutaneous, 3-day, continuous glucose-sensing system for three consecutive 72-h periods. Subjects were randomly assigned (1:1 ratio) to either a control group (continuous glucose data not provided) or a display group (continuous glucose data not provided during period 1 but displayed during periods 2 and 3). During periods 2 and 3, patients in the display group had real-time access to sensor glucose values, could review glucose trends over the preceding 1, 3, and 9 h, and were provided with high (>= 200 mg/dl) and low (