Absolute calibration and monitoring of a spectrometric channel using a photomultiplier
Absolute calibration and monitoring of a spectrometric channel using a photomultiplier
复制标题
DOI:
10.1016/0168-9002(94)90183-x
复制
发表时间:
1994-02
影响因子:
1.4
通讯作者:
E. Bellamy;G. Bellettini;F. Gervelli;M. Incagli;D. Lucchesi;C. Pagliarone;F. Zetti;Yury A. Budagov-Yury-A.
中科院分区:
文献类型:
--
作者:
E. Bellamy;G. Bellettini;F. Gervelli;M. Incagli;D. Lucchesi;C. Pagliarone;F. Zetti;Yury A. Budagov-Yury-A.
Scintillation detectors (counters, calorimeters etc.) are presently extensively used and will be used in future experiments at new accelerators. Usually in such detectors photomultiplier tubes (PM) are used for the detection of light. In detectors of this type there exists not only an intrinsic spread in characteristic parameters among different PMs, but also some time dependence of parameters of a given PM. Therefore a system of calibration and monitoring of the PM based spectrometric channels is an important part of the experimental setups employing scintillation detectors. Particular interest is paid to the absolute calibration, ie to the measurement of the energy released in the scintillators in terms of the number of photoelectrons emitted from the photocathode. Usually the performances of detectors using scintillation light are qualified by this ratio, ie in units of photoelectrons per GeV.