Wear rate and osteolysis in two types of second-generation annealed highly cross-linked polyethylene in total hip arthroplasty: A retrospective comparative study with a minimum of five years
Wear rate and osteolysis in two types of second-generation annealed highly cross-linked polyethylene in total hip arthroplasty: A retrospective comparative study with a minimum of five years
复制标题
两种第二代退火高交联聚乙烯在全髋关节置换术中的磨损率和骨质溶解:至少五年的回顾性比较研究
DOI:
10.1016/j.otsr.2021.103147
复制
发表时间:
2021
期刊:
影响因子:
--
通讯作者:
Okawa Atsushi
中科院分区:
文献类型:
--
作者:
Yamamoto Naoki;Takada Ryohei;Jinno Tetsuya;Miyatake Kazumasa;Watanabe Naoto;Koga Hideyuki;Yoshii Toshitaka;Yagishita Kazuyoshi;Okawa Atsushi
BackgroundAs no previous study has directly compared the linear wear rate in two types of second-generation annealed highly cross-linked polyethylene, we performed a retrospective study with a minimum of 5-year follow-up to assess primary arthroplasties in the (1) wear rates and (2) incidence of osteolysis of the two types of HXLPE.HypothesisThere was no significant difference in the linear wear rate and the incidence of osteolysis between the two types of second-generation annealed highly cross-linked polyethylene.Patients and methodsIn this single-center study, we reviewed 257 cases of primary cementless total hip arthroplasties between 2011 and 2015, which were performed with 32 mm delta ceramic on second-generation annealed highly cross-linked polyethylene (X3 and E1 were used in 105 and 103 cases, respectively.). The mean wear rate was evaluated using a computer-assisted method, and the incidence of osteolysis was evaluated based on the appearance of a localized area with loss of trabecular bone or cortical erosion adjacent to the implants during the latest follow-up.ResultsIn total, we evaluated 208 cases, followed postoperatively for over 5 years (mean, 6.1 years, range: 5.0-8.0). There were no significant differences between the two groups with respect to age (list in order of Group X, Group E,pvalue) (61.2 ± 12.3, 62.7 ± 12.1,p= 0.36), sex (ratio of male: 17.1%, 14.6%,p= 0.61), body mass index (22.9 ± 3.7, 22.8 ± 4.0,p= 0.91), pre- (49.9 ± 14.8, 48.5 ± 13.8,p= 0.49) and post-operative (91.3 ± 9.1, 92.7 ± 7.0,p= 0.23) Japanese Orthopaedic Association Hip Score, cup size (50.8 ± 3.0, 50.9 ± 2.2,p= 0.70), cup inclination (38.7 ± 4.8, 37.6 ± 4.8,p= 0.10), and cup anteversion (18.7 ± 6.9, 18.5 ± 7.6,p= 0.80). The mean linear wear rates of the X3 and E1 groups were 0.057 ± 0.039 (range: 0–0.16) and 0.054 ± 0.037 mm/year (range: 0–0.15), respectively (p= 0.61). No osteolysis was found on the final plain radiographs in both groups.DiscussionThis study revealed that both types of highly cross-linked polyethylene have excellent linear wear rates and were equally safe to use. However, the difference between the two materials in terms of the long-term wear rate should be further validated.Level of evidenceIII; retrospective case control study.