An acoustic plethysmograph to measure total infant body volume.
An acoustic plethysmograph to measure total infant body volume.
复制标题
用于测量婴儿身体总体积的声学体积描记器。
DOI:
10.1115/1.3138560
复制
发表时间:
1985
期刊:
影响因子:
--
通讯作者:
Garza,C
中科院分区:
文献类型:
--
作者:
Deskins,WG;Winter,DC;Sheng,HP;Garza,C
An acoustic plethysmograph intended to measure the body volume of premature infants has been developed using the principle of the Helmholtz resonator, in which the resonance frequency is dependent on the volume of the resonating cavity. A prototype system was built and used to measure the volume of inanimate objects and newborn miniature pigs. Results for inanimate objects agree within 1 percent with comparable measurements by water displacement. Results of the animal body volume measurements compare favorably (within an average of 1.1 percent) with those obtained using hydrostatic weighing.