Experimental assessment of human corneal hysteresis
Experimental assessment of human corneal hysteresis
复制标题
DOI:
10.1080/02713680701882519
复制
发表时间:
2008-03-01
影响因子:
2
通讯作者:
Garway-Heath, David
中科院分区:
文献类型:
--
作者:
Elsheikh, Ahmed;Wang, Defu;Garway-Heath, David
Purpose: Hysteresis is a viscoelastic property characterized by the difference in behavior under loading and unloading. The aim of the study was to determine corneal hysteresis using experimental means. Methods: Twenty-nine human corneas with 50-95 year age were subjected to cycles of pressure loading and unloading. Two pressure application rates were adopted to approximate static and dynamic loading conditions. Results: The behavior under both loading and unloading was found to stiffen with increased age. The unloading behavior appeared to be largely independent of the pressure level at which unloading started. The difference between the behavior patterns under loading and unloading was quantified and used as a measure of corneal hysteresis. The hysteresis area was significantly larger with faster loading and with decreased age. Conclusions: The trend for hysteresis to decrease with age is in agreement with previous clinical observations. Hysteresis was also found to increase with faster pressure application.