Repeated immunosensing by a dithiobis(succinimidyl propionate)-modified SAW device
Repeated immunosensing by a dithiobis(succinimidyl propionate)-modified SAW device
复制标题
通过二硫双(琥珀酰亚胺基丙酸酯)修饰的 SAW 装置进行重复免疫传感
DOI:
10.1016/j.talanta.2019.05.080
复制
发表时间:
2019
期刊:
影响因子:
6.1
通讯作者:
Mitsubayashi Kohji
中科院分区:
文献类型:
--
作者:
Toma Koji;Oishi Koki;Yoshimura Naoyuki;Arakawa Takahiro;Yatsuda Hiromi;Mitsubayashi Kohji
The possibility of dithiobis(succinimidyl propionate) (DSP) for repeated immunoassay with a surface acoustic wave (SAW) immunosensor was explored. In the sensor, DSP was used to modify a gold-coated quartz sensing area of a SAW device by forming a self-assembled monolayer on the gold surface. In a model sandwich assay using mouse (mIgG) and anti-mouse (a-mIgG) antibodies, the primary antibody, mIgG, firstly reacted with N-hydroxysuccinimide ester groups of DSP and was immobilized on the SAW device to fabricate the SAW immunosensor. Optimization of adsorption time of mIgG revealed that both degrees of adsorption and immobilization of mIgG reached a saturation at 30 min although the immobilization was more dependent on the adsorption time. Through characterization of the DSP-modified SAW immunosensor, a high selectivity, with which no sensor output was observed from various kinds of secondary antibodies except for a-mIgG, along with 8-fold shorter measurement time (15 min) than that of enzyme-linked immunosorbent assay were obtained. Furthermore, 10 repeated measurement of a-mIgG demonstrated a high reproducibility of the sensor output (coefficient of variation of 7.0%). These validate the utility of DSP in the SAW immunosensor for rapid and repeated measurement of antigens.