GLUCOSE ENZYME ELECTRODE WITH EXTENDED LINEARITY - APPLICATION TO UNDILUTED BLOOD MEASUREMENTS
GLUCOSE ENZYME ELECTRODE WITH EXTENDED LINEARITY - APPLICATION TO UNDILUTED BLOOD MEASUREMENTS
复制标题
DOI:
10.1016/0003-2670(86)80074-5
复制
发表时间:
1986-05-31
影响因子:
6.2
通讯作者:
VADGAMA, PM
中科院分区:
文献类型:
--
作者:
MULLEN, WH;KEEDY, FH;VADGAMA, PM
An enzyme electrodes is described for glucose determination in unstirred, undiluted whole blood. The system comprises an H2O2-detecting electrode upon which is placed a membrane laminate incorporating glucose oxidase. The external membrane was pretreated with methyltrichlorosilane. The electrode response was linearly dependent on glucose concentration up to 50 mmol l−1glucose, it had a decreased dependence on dissolved oxygen concentrations and gave response times of 30–90 s. Whole blood glucose measurements correlated well with a routine spectrophotometric method.