Comparison of tissue oximeters on a liquid phantom with adjustable optical properties
Comparison of tissue oximeters on a liquid phantom with adjustable optical properties
复制标题
DOI:
10.1364/boe.7.002973
复制
发表时间:
2016-08-01
影响因子:
3.4
通讯作者:
Wolf, M.
中科院分区:
文献类型:
--
作者:
Kleiser, S.;Nasseri, N.;Wolf, M.
The SafeBoosC trial showed that cerebral oximetry combined with a treatment guideline can reduce the the burden of hypoxia in neonates by 50% [Brit. Med. J. 350, g7635 (2015)]. However, guidelines based on oximetry by one oximeter are not directly usable by other oximeters. We made a blood-lipid phantom simulating the neonatal head to determine the relation between oxygenation values obtained by different oximeters. We calculated coefficients for easy conversion from one oximeter to the other. We additionally determined the corresponding SafeBoosC intervention thresholds at which we measured an uncertainty of up to 9.2% when varying hemoglobin content from 25 mu M to 70 mu M. In conclusion, this paper makes the comparison of absolute values obtained by different oximeters possible. (C) 2016 Optical Society of America