Field comparison of commercially available short-term radon detectors

Field comparison of commercially available short-term radon detectors
复制标题

DOI:
10.1097/01.hp.0000206040.55298.ce
复制
发表时间:
2006-09-01
期刊:
影响因子:
2.2
通讯作者:
Field, R. William
Field, R. William
中科院分区:
医学4区
文献类型:
--
作者:
Sun, Kainan;Majdan, Marek;Field, R. William

文献摘要

被引文献

相似文献

我们对市售的短期氡探测器进行了比较,以确定这些探测器在实际野外条件下的准确度和精密度。在野外条件下,我们让来自六个连队的15个测氡仪暴露在一个参考的氡浓度下。在接受测试的六家公司中,有五家没有通过美国环境保护局之前制定的准确性准则(均为个人相对误差
We performed a comparison of commercially available short-term radon detectors in order to determine the accuracy and precision of the detectors under actual field conditions. We exposed fifteen radon detectors, under field conditions, from each of six companies to a reference radon concentration. Five of the six companies tested did not pass the U.S. Environmental Protection Agency's previously established accuracy guideline (all individual relative errors