Use of a modified walkley-black method to determine the organic and elemental carbon content of urban aerosols collected on glass fibre filters
Use of a modified walkley-black method to determine the organic and elemental carbon content of urban aerosols collected on glass fibre filters
复制标题
使用改进的沃克利黑法测定玻璃纤维过滤器收集的城市气溶胶的有机碳和元素碳含量
DOI:
10.1016/0045-6535(95)00308-u
复制
发表时间:
1995
期刊:
影响因子:
8.8
通讯作者:
G. Bailey
中科院分区:
文献类型:
--
作者:
Y. Chan;P. Vowles;G. McTainsh;R. Simpson;D. Cohen;G. Bailey
An analysis based on the Walkley-Black method is a rapid and effective means for determining the organic carbon, elemental carbon and total organic carbon content of urban aerosol samples collected on glass fibre and teflon-backed glass fibre filters. The estimated accuracy of the method is + 18%, −33% and −8% for OC, EC and TOC, respectively, the precision is ±7%, ±21% and ±4%, and the detection limit 0.33 μg m−3, 0.54 μg m−3and 0.29 μg m−3.