Tritium digital autoradiography with a Medipix2 hybrid silicon pixel detector
Tritium digital autoradiography with a Medipix2 hybrid silicon pixel detector
复制标题
使用 Medipix2 混合硅像素探测器进行氚数字放射自显影
DOI:
10.1016/j.nima.2003.09.027
复制
发表时间:
2004
影响因子:
1.4
通讯作者:
P. Russo
中科院分区:
文献类型:
--
作者:
G. Mettivier;M. Montesi;P. Russo
We report on the first measurements of3H beta autoradiography obtained using a room temperature hybrid pixel detector, consisting of the Medipix2 single particle counting read-out chip bump-bonded to a 300μm thick silicon pixel detector. This system has 256×256 square pixels of 55μm pitch for a total sensitive area of 14×14mm2. Each pixel contains a double threshold discriminator and a 13-bit counter. Using a detection threshold equivalent to less than 6keV and a background count rate of 5×10−3counts mm−2s−1, with exposures up to several hours, real-time images have been obtained of tritium-labeled solution drops spotted on a thin mylar foil placed in contact with the continuous top electrode of the silicon detector, in open air condition.