A microcalorimeter EDS system suitable for low acceleration voltage analysis
A microcalorimeter EDS system suitable for low acceleration voltage analysis
复制标题
适合低加速电压分析的微量热仪EDS系统
DOI:
10.1002/sia.2408
复制
发表时间:
2006
影响因子:
1.7
通讯作者:
S. Nakayama
中科院分区:
文献类型:
--
作者:
Keiichi Tanaka;A. Odawara;A. Nagata;Masanori Ikeda;Yukari Baba;S. Nakayama
When a scanning electron microscope (SEM) is used for analysis, a low acceleration voltage must be used to reduce damage to the sample and to obtain information near the sample surface. Here, a microcalorimeter energy dispersive spectrometer (EDS) system without external cryogen was developed to analyze material samples under low acceleration voltages (<5 kV) with high energy resolution (<20 eV). This EDS system can maintain the operating temperature of the transition edge X‐ray sensor (TES) composing of the microcalorimeter system continuously at 100 mK to measure the X‐ray spectrum with high reproducibility and with high energy resolution. This system resembles a conventional EDS system, but also has a snout to allow the X‐ray detector to be close to the sample. The measured energy resolution was 19 eV (5.9 keV) and the count rate was 300 cps. A thin X‐ray window of aluminum/polyimide bilayer was designed to transmit the X rays at a low acceleration voltage (5 keV). When SUS‐304 was analyzed using this EDS system at an acceleration voltage less than 3 kV, the Cr Lα (573 eV) was distinguished between the closely spaced O Kα (525 eV) and Fe Ll (615 eV)lines, whereas when a solid‐state detector(SSD) was used, it was not distinguished owing to inadequate energy resolution (130 eV). Copyright © 2006 John Wiley & Sons, Ltd.