Spatial resolution of a laboratory based X-Ray cone-beam laminography scanning system for various trajectories
Spatial resolution of a laboratory based X-Ray cone-beam laminography scanning system for various trajectories
复制标题
DOI:
10.1016/j.ndteint.2020.102222
复制
发表时间:
2020-04-01
影响因子:
4.2
通讯作者:
Blumensath, Thomas
中科院分区:
文献类型:
--
作者:
Deyhle, Hans;Towsyfyan, Hossein;Blumensath, Thomas
Computed laminography (CL), a long established NDT method, is ideal for the non-destructive evaluation of plate-like structures, for which conventional computed tomography (CT) is less well suited. This paper demonstrates the feasibility of implementing raster CL on a Nikon Metrology custom build X-ray CT scanner without the need for specialised equipment. To provide a more principled way to demonstrate the widespread adoption of CL for defect characterization in plate-like structures, a test specimen was designed and manufactured from Polyoxymethylene (Delrin) that allows well defined features of different size and orientation to be characterized. The study highlights the relation between direction-dependent spatial resolution, feature geometry and scan trajectory. It demonstrates that high-resolution and high-speed in situ volume-imaging of plate-like structures with large aspect ratios is viable at levels that would be challenging via CT or alternative three-dimensional NDT methods.