Effect of Continuous Cuff Pressure Regulator in General Anaesthesia with Laryngeal Mask Airway
Effect of Continuous Cuff Pressure Regulator in General Anaesthesia with Laryngeal Mask Airway
复制标题
连续袖带压力调节器在喉罩气道全身麻醉中的作用
DOI:
10.1177/147323001103900534
复制
发表时间:
2011
影响因子:
1.6
通讯作者:
J
中科院分区:
文献类型:
--
作者:
Y. Jeon;J.W. Choi;H. Jung;Y. Kim;D. Kim;J. Kim;J
Postoperative pharyngolaryngeal complications (PPLC) occur during anaesthesia due to increased cuff pressure following the insertion of laryngeal mask airways. The use of a pressure regulator to prevent PPLC was evaluated in a prospective, randomized study. Sixty patients scheduled to receive general anaesthesia were randomly assigned to two equal groups of 30, either with or without the regulator. The ‘just seal’ cuff pressure (JSCP), cuff pressure at 5-min intervals during anaesthesia, incidence of pharyngeal sore throat (PST), dysphagia, dysphonia and other complications were evaluated at 1 and 24 h postoperatively. The combined mean ± SD JSCP of both groups was 20.3 ± 3.2 mmHg. In the group with the regulator, cuff pressure was maintained at a constant level during anaesthesia. This study demonstrated that the regulator is a simple, functional device that can reduce the incidence of PST significantly at 1 h postoperatively, following general anaesthesia.