Counting single chromophore molecules for ultrasensitive analysis and separations on microchip devices
Counting single chromophore molecules for ultrasensitive analysis and separations on microchip devices
复制标题
DOI:
10.1021/ac9707242
复制
发表时间:
1998-02-01
影响因子:
7.4
通讯作者:
Ramsey, JM
中科院分区:
文献类型:
--
作者:
Fister, JC;Jacobson, SC;Ramsey, JM
Separations of 15 pM rhodamine 6G and 30 pM rhodamine B performed in a micromachined electrophoresis channel were detected by counting fluorescence bursts from individual molecules. The migration times, peak widths, and analyte concentrations were estimated from the number and the migration time distribution of the detected molecules. Concentration detection limits estimated at >99% confidence were 1.7 pM rhodamine 6G and 8.5 pM rhodamine B. The separations required