Accurate measurement of three-dimensional knee replacement kinematics using single-plane fluoroscopy
Accurate measurement of three-dimensional knee replacement kinematics using single-plane fluoroscopy
复制标题
DOI:
10.1109/10.495283
复制
发表时间:
1996-06-01
影响因子:
4.6
通讯作者:
Hodge, WA
中科院分区:
文献类型:
--
作者:
Banks, SA;Hodge, WA
A simple extension of a previously reported object recognition technique has been used to implement a six-degree-of-freedom position/orientation estimator for the measurement of knee replacement motion from two-dimensional (2-D) fluoroscopic images. Computer modeling studies and controlled mechanical tests were performed to assess the accuracy of the technique. The results indicate that knee rotations can be measured with an accuracy of approximately one degree and that sagittal plane translations can be measured with an accuracy of approximately 0.5 mm. The measurement technique is uniquely well suited for performing dynamic kinematic measurements on individuals with knee replacements, and for performing comparative studies among subjects with different designs of knee replacements.