Safety-assured development of the GPCA infusion pump software
Safety-assured development of the GPCA infusion pump software
复制标题
安全可靠的 GPCA 输液泵软件开发
DOI:
--
复制
发表时间:
2011
期刊:
影响因子:
--
通讯作者:
R. Jetley
中科院分区:
文献类型:
--
作者:
Baekgyu Kim;A. Ayoub;O. Sokolsky;Insup Lee;P. Jones;Y. Zhang;R. Jetley
This paper presents our effort of using model-driven engineering to establish a safety-assured implementation of Patient-Controlled Analgesic (PCA) infusion pump software based on the generic PCA reference model provided by the U.S. Food and Drug Administration (FDA). The reference model was first translated into a network of timed automata using the UPPAAL tool. Its safety properties were then assured according to the set of generic safety requirements also provided by the FDA. Once the safety of the reference model was established, we applied the TIMES tool to automatically generate platform-independent code as its preliminary implementation. The code was then equipped with auxiliary facilities to interface with pump hardware and deployed onto a real PCA pump. Experiments show that the code worked correctly and effectively with the real pump. To assure that the code does not introduce any violation of the safety requirements, we also developed a testbed to check the consistency between the reference model and the code through conformance testing. Challenges encountered and lessons learned during our work are also discussed in this paper.