Failure of a Heifetz aneurysm clip.

Failure of a Heifetz aneurysm clip.
复制标题

海菲茨动脉瘤夹失败。

DOI:
10.3171/jns.1982.57.2.0233
复制
发表时间:
1982
影响因子:
4.1
通讯作者:
Y. Kéravel
Y. Kéravel
中科院分区:
医学1区
文献类型:
--
作者:
R. Kossowsky;M. Dujovny;N. Kossovsky;Y. Kéravel

文献摘要

被引文献

相似文献

一名16岁女孩死于海菲兹动脉瘤夹刀片断裂后的急性蛛网膜下腔出血。夹子由17- 7 PH钢制成,在冶金测试中发现其对晶间腐蚀高度敏感。断裂机制是应力腐蚀,由应力载荷、电解环境和敏感钢的组合引起。
A 16-year-old girl died from an acute subarachnoid hemorrhage following the fracture of a blade of a Heifetz aneurysm clip. The clip was manufactured from 17-7PH steel, which on metallurgical testing was found to be highly sensitive to intergranular corrosion. The fracture mechanism was stress corrosion, brought on by the combination of a stress load, an electrolytic environment, and a susceptible steel.