Uveal and capsular biocompatibility of an intraocular lens with a hydrophilic anterior surface and a hydrophobic posterior surface
Uveal and capsular biocompatibility of an intraocular lens with a hydrophilic anterior surface and a hydrophobic posterior surface
复制标题
具有亲水性前表面和疏水性后表面的人工晶状体的葡萄膜和囊生物相容性
DOI:
10.1016/j.jcrs.2009.09.027
复制
发表时间:
2010-02-01
影响因子:
2.8
通讯作者:
Wang, Yao
中科院分区:
文献类型:
--
作者:
Huang, Xiao-Dan;Yao, Ke;Wang, Yao
PURPOSE: To evaluate the uveal and capsular biocompatibility of intraocular lenses (IOLs) with a hydrophilic anterior surface and a hydrophobic posterior surface in a rabbit model.SETTING: Eye Center, Affiliated Second Hospital, College of Medicine, Zhejiang University, Hangzhou, China.METHODS: Modified silicone IOLs were produced by grafting 2-methacryloyloxyethyl phosphorylcholine (MPC) onto the anterior IOL surface using a plasma technique. A contact-angle test characterized the hydrophilicity of the IOL surface; physical and optical properties were determined by State Food and Drug Administration (SFDA) standards. Rabbits had phacomulsification and implantation a modified silicone IOL, a control silicone IOL, or a hydrogel IOL. Postoperative inflammation was assessed by aqueous flare measurement, and PCO was evaluated by software analysis. Three months after surgery, attached cells on extracted IOLs were evaluated by light microscopy; PCO was evaluated by Miyake-Apple technique. Histologic sections of globes were used to assess lens epithelial cells (LECs) and extracellular matrix in the capsular bag.RESULTS: Contact angle data showed the MPC-modified IOL had a hydrophilic anterior surface and hydrophobic posterior surface. The properties of the modified IOLs met SFDA standards. There was no statistical difference in aqueous flare between the IOL groups at any time. The modified and control IOLs had less PCO than the hydrogel IOLs (P