Use of diffusing filters for artificially reducing visual acuity when testing equipment and procedures
Use of diffusing filters for artificially reducing visual acuity when testing equipment and procedures
复制标题
DOI:
10.1007/s10633-019-09715-5
复制
发表时间:
2020-02-01
影响因子:
1.4
通讯作者:
Struebin, Isabell
中科院分区:
文献类型:
--
作者:
Heinrich, Sven P.;Struebin, Isabell
Purpose When evaluating ophthalmological devices and procedures, for instance those for visual electrophysiology, it is often desirable to perform tests with reduced acuity. Doing this with individuals with actual visual impairments has a number of disadvantages, such as considerable recruitment efforts, especially when a specific acuity range is targeted, and little control about the actual perceptual characteristics of the impairment, which are normally not fully known. Lenses with positive diopters or blurring filters that are placed in front of the eyes of visually normal observers promise a simple solution to the problem. However, defocus results in considerable spurious resolution, and previous studies suggest that the frequently used Bangerter occluders are not optimal for the purpose. The present study therefore reviews a number of other options and tests a selection of filters with respect to their effect on acuity and contrast sensitivity with the aim of identifying filters that primarily degrade acuity while mostly sparing contrast sensitivity. Methods First, we screened several filters for potential usefulness. The Freiburg Acuity and Contrast Test was then used to measure visual acuity and contrast sensitivity with a subset of three filters (Luminit LSD 0.5 degrees and 1 degrees, and LEE 420) and, for comparison, with a Bangerter occluder with a nominal acuity grade of 0.1. A qualitative comparison of the filters' effect on the checkerboard-reversal VEP was also performed. Results With both Luminit filters, variability in acuity across participants was relatively small, and at least with the 0.5 degrees version, contrast sensitivity was relativity little affected. The LEE filter and the Bangerter occluder resulted in more variability and, compared to the effect on acuity, a relatively strong reduction in contrast sensitivity. Comparing the Luminit 0.5 degrees and 1 degrees filters, the reduction of acuity was not proportional to physical stimulus degradation. The effect on VEP responses was consistent with the psychophysical data. Conclusions The Luminit filters, which have a Gaussian light diffusion profile, appear to be a good choice for artificial reduction of acuity.